Biosafety monitoring systems

Enviro’s ETS-BSS-GXX is a pioneering BSS Monitoring and Alarm System developed to ensure the safety of users working in pharmaceutical and biological safety cabinets. Boasting a simple, yet intuitive, user interface that has been optimized for user experience, the ETS-BSS-GXX Series is easy to navigate and features simple on-screen prompts to guide end users.
ETS-BSS-GXX unit is a Microcontroller Electronic BSS system that consists of a front panel sticker with a 128X64 Graphical Display, Mainboard, and Relay board. The front (Human interface board) panel consists of 5 Nos. LEDs ALARM, BLOWER, LIGHT, SWITCH, UV, and four Programmable keys MENU, UP, DOWN, SET/DIAGNOSTIC. A total of 8 programming Keys are used for programming different parameters and operating LIGHT, BLOWER, SWITCH, and UV.

CE Certified

BIS Certified

Make in India

Description
  • Advanced Micro-controller based.
  • 128 x 64 G-LCD display.
  • Four Soft Touch: ON/OFF Action Key for Blower, UV, Lamp & Socket.
  • 3 Digital Reed Input for SASH window status indication.
  • Inbuilt UV hour meter with reset option.
  • User-programmable system with password-protected menu.
  • Blower Air-Flow Monitoring in m/s or fpm for inlet & outlet blower.
  • Temperature, Inlet/Outlet Airflow Monitoring.
  • PC connectivity with RS-485 output for system status.
Technical Information

Specifications

Description

Type

Microcontroller Based Electronic Enviro Bio Safety System

Range

0-1m/s (depends on sensor) 0-5V output for 0-1m/s & 0 to 99.9 Deg C

Control resolution

0.01 m/sec ; 0.1 Deg C

Relay Output

4 Relay output Blower,Light,Socket, UV

COM port

RS 485 with Periodic System Status

SASH Alarm

Yes with alarm repeat

Horn Silence

Yes

Display & LED

128 X 64 Graphical LCD Display

Led Indication => ALARM, BLOWER, LIGHT, SWITCH, UV

Keypad

Total 8 Keys

Menu Keys =>SEL, INC, DEC, SET/DIAGNOSTIC

Function Keys => BLOWER, LIGHT, SWITCH, UV

Analog I/Ps

Total 3 Analog I/Ps

1. 1st Channel:- RTD PT100 Sensor (0.0°C to 99.9°C)

2. 2nd Channel:-Air Flow Sensor (0 to 5VDC) Range 0 to 1m/s

3. 3rd Channel:-Air Flow Sensor (0 to 5VDC) Range 0 to 1m/s

Digital Inputs

Total 3 Magnetic Reed Switch Inputs

1. Input1 : Reed Switch (CLOSED)

2. Input2 : Reed Switch (READY TO USE)

3. Input2 : Reed Switch (OPEN)

Potential Free Outputs

Total 4 Potential Free Outputs With Contact Rating Of 5Amp Protected By 5Amp Fuse.

1. Relay1: BLOWER

2. Relay2: LIGHT

3. Relay3: SWITCH

4. Relay4: UV

Power Supply

1) 12VDC @ 1Amp/50Hz For Mother Board

2) 12VDC @ 1Amp/50Hz For Relay Board

Mounting

Fully Flush