enviroDPMS LCD
RH+T+DP CLEAN ROOM MONITOR
Assists in GMP and Data Integrity in multiple pharma clean rooms.
CE Certified
BIS Certified

Make in India
- Microcontroller based.
- RH, Temperature & DP measurement.
- 3.5-inch color LCD/TFT display.
- Software programmable channel ranges, units & input types.
- User programmable for 1, 2, or 3-channel measurement.
- Real-Time Clock with battery backup.
- Software calibration for channels.
- RS485 Communication with MODBUS / TCP multidrop protocol for PLC, SCADA, etc.
- 24 VDC, 1 AMP Power Supply.
- FDA 21 CFR Part 11 Compliance EMS software.
- Advanced digital RH+T sensor technology.
|
Parameter |
Description |
|---|---|
|
No of Input |
3 (RH, T & DP) |
|
Input Range |
Humidity: 0-100% RH, Temperature PT-100: -90°C ~ 390°C (user selectable); Digital: 0°C ~ 100°C (user selectable), Diff Pressure: -250 ~ 900Pa; -25 ~ 90 mmWC (user selectable) |
|
Accuracy |
(For In-built Sensors) : Humidity: ± 2% RH, Temperature: ± 0.3°C, Diff Pressure: ± 1.5% of reading ± 5 Pa; ± 1.5% of reading ± 0.5 mmWC (For External Sensors) : Humidity: ± 2% RH (depending on sensor), Temperature: ± 0.5% (depending on sensor), Diff Pressure: 0.5% |
|
Display |
3.5 inch color LCD/TFT display |
|
Buzzer Output |
External Potential free Relay output for Buzzer |
|
Clock |
Real-time clock: Available, Battery backup: Available (Optional) |
|
Dimensions |
Front Plate: 210(W)x210(H) mm Back Plate: 190(W)x150(H)x32(D) mm |
|
Enclosure |
Faceplate: SS304 Body: MS, powder-coated |
|
Communication Port |
2 Nos RS485 (1st RS485 for PC Comm./MODBUS & 2nd for External Display) |
|
Resolution |
0.1 for RH & T 0.1 for DP unit mmWC 1 for DP unit Pa |
|
Supply Voltage |
24 V DC, 1 amp |
|
Operating Ambient Temperature |
0–50°C |
|
Alarm & Alerts |
Alarms & Alerts by way of Pop-ups, Email & SMS can be sent to multiple authorities |