enviroDPMS LCD FLP
RH+T+DP CLEAN ROOM MONITOR
Assists in GMP and Data Integrity in multiple pharma clean rooms
CE Certified
BIS Certified
Make in India
Description
- Microcontroller based.
- RH, Temperature & DP measurement (Flameproof).
- 3.5-inch color LCD/TFT display.
- Software Programmable channel ranges, unit & input types.
- User programmable for 1,2 or 3-channel measurement.
- Real-Time Clock with battery backup.
- Software calibration for channels.
- MODBUS multidrop communication for PLC, SCADA, etc.
- 24 VDC, 1 AMP Power Supply.
- FDA 21 CFR Part 11 Compliance EMS software.
- Advanced digital RH+T sensor technology.
Technical Information
Parameter |
Description |
---|---|
No. of inputs |
3 (RH, T & DP) |
Input Range |
Humidity- 0~100% RH Temperature PT-100: -90°C ~ 390°C (user selectable) Digital: 0°C ~ 100°C (user selectable) Diff Pressure: -250 ~ 900Pa; -25 ~ 90 mmWC (user selectable) |
Accuracy |
(For In-built Sensors): Humidity - ± 2% RH, Temperature - ± 0.3 °C, Diff Pressure - ± 1.5% of reading ± 5 Pa; ± 1.5% of reading ± 0.5 mmWC (For External Sensors): Humidity -± 2% RH (depending on sensor), Temperature - ± 0.5% (depending on sensor), Diff Pressure - 0.5% (depending on sensor) |
Display |
3.5 inch color LCD/TFT display |
Buzzer output |
External Potential free relay output for Buzzer |
Clock |
Real time clock: Available Battery backup: Available (Optional) |
Dimensions |
Without Plate - 210(W)x210(H)x50(D) With Plate - 260(W)x250(H)x60(D) mm |
Enclosure |
Die Cast Aluminium Alloy LM-6 SS304 Faceplate (optional) CertFlameproof for Gas Groups, IIA/IIB, Zone 1&2, IP55 Protection |
Communication Port |
2 Nos RS485 (1st RS485 for PC Comm./MODBUS & 2nd for External Display) |
Resolution |
0.1 for RH & T 0.1 for DP unit mmWC 1 for DP unit Pa |
Supply voltage |
24 V DC, 1 amp |
Operating ambient temperature |
0~50 °C |
Alarm & Alerts |
Alarms & Alerts by way of Pop-ups, Email & SMS can be sent to multiple authorities |